Overview
Input 115 V AC, 230 V AC
Output 230 V AC/3.05 A
Power 700 VA
No. of outputs: 6 x
Equipotential bonding plug according to DIN 42801
Inrush current limitation by electronic start-up current limitation with half-shaft failure detection
Input 115 V AC, 230 V AC
Output 230 V AC/3.05 A
Power 700 VA
No. of outputs: 6 x
Equipotential bonding plug according to DIN 42801
Inrush current limitation by electronic start-up current limitation with half-shaft failure detection
According to the EU Directive 93/42/EEC must medical electrical (ME) devices and systems be designed so that patients, operators and other persons not to harm. Through the use of a medical isolating transformer PolyMIT of sedlbauer is the contact current from a connected device or multiple devices to values far below the defined limit values in accordance with the EN 60601-1, 3rd Edition is limited. The transformer is TÜV certified in accordance with the EN 60601-1-1, 3rd edition. The transformers will impress you with the highest energy efficiency (efficiency up to 98%) in the tightest of spaces and have a very low magnetic field. Short-circuit and overload protection by fuse and reversible temperature switch.
Inrush current limitation by electronic start-up current limitation with half-shaft failure detection · Short-circuit and overload protection by fuse and reversible temperature switch.
Power cable, instructions for use, safety catch, spare fuse.